MDCG 2021-24 explained
How to classify a medical device under EU MDR Annex VIII, every change in MDCG 2021-24 Rev.1 (April 2026), Rule 11 for software, and what the 2025 proposal would change.
MDR IVDR qualification, settled from the intended purpose before any class, decides which EU regulation applies. A product meant to examine human specimens in vitro for medical information is an IVD under the IVDR, Regulation (EU) 2017/746. Other medical devices fall under the MDR, Regulation (EU) 2017/745. A CE mark under the wrong one shows no conformity.
The IVDR is the EU regulation for in vitro diagnostic medical devices. IVDR rules have applied since 26 May 2022, when they replaced Directive 98/79/EC (IVDR, Articles 112 and 113).
To decide, read everything you say about the product against the IVD definition in IVDR Article 2(2) and the device definition in MDR Article 2(1), check the exclusions and the integral-product rule, and only then classify.
| Milestone | Date | Applies to | Source |
|---|---|---|---|
| MDR applies | 26 May 2021 | medical devices | European Commission, overview |
| IVDR applies | 26 May 2022 | all IVDs, with no transition for new IVDs or non-sterile class A | IVDR, Article 113(2) and Commission Q&A, July 2024 |
| IVDR quality management system in place | 26 May 2025 (passed) | legacy IVDs using the transition | Regulation (EU) 2024/1860 and Commission Q&A |
| Notified body application, then signed agreement | 26 May 2025, then 26 September 2025 (both passed) | IVDD-certified legacy devices and class D | Commission Q&A |
| Notified body application, then signed agreement | 26 May 2026, then 26 September 2026 (both passed) | class C legacy devices | Commission Q&A |
| Notified body application, then signed agreement | 26 May 2027, then 26 September 2027 | class B and sterile class A legacy devices | Commission Q&A |
| End of the transitional period | 31 December 2027 / 31 December 2028 / 31 December 2029 | IVDD-certified and class D / class C / class B and sterile class A | Commission Q&A |
| Commission proposal COM(2025)1023 to simplify MDR and IVDR | proposed 16 December 2025, awaiting committee decision, indicative plenary 18 January 2027 | not law | European Parliament, procedure 2025/0404(COD) |
IVDR is the In Vitro Diagnostic Medical Devices Regulation, Regulation (EU) 2017/746. It sets the rules for placing in vitro diagnostic devices on the EU market and has applied since 26 May 2022, when it replaced the In Vitro Diagnostic Medical Devices Directive 98/79/EC (IVDD). An IVD is the product, and the IVDR is the law that governs it.
Under IVDR Article 2(2), an IVD is a medical device that is a reagent, calibrator, control material, kit, instrument, apparatus, piece of equipment, software or system, intended by its manufacturer to be used in vitro to examine specimens derived from the human body, including blood and tissue donations. Its purpose, solely or principally, is information on a physiological or pathological process or state, on congenital impairments, on predisposition to a disease, on safety and compatibility with potential recipients, on likely treatment response, or for defining or monitoring therapeutic measures. Specimen receptacles count as IVDs as well.
IVDR Article 1(3) takes four groups out of scope. The first is products for general laboratory use and research-use-only products, unless the manufacturer specifically intends them for in vitro diagnostic examination. The second is invasive sampling products and products applied directly to the body to obtain a specimen. The last two are internationally certified reference materials and materials for external quality assessment schemes.
An IVD placed on the market through the IVDR conformity assessment carries the CE marking (IVDR, Article 18), followed by the notified body's number where one took part. "CE-IVD" is informal shorthand for a CE-marked IVD, not a separate approval, and it does not tell you the route on its own. Legacy IVDs still in their transitional period carry CE marking based on IVDD conformity, and the EU declaration of conformity says which regulation it rests on.
The MDR covers medical devices in general, including diagnostic devices that work without a specimen. The IVDR covers the medical devices that examine human specimens in vitro. The MDR does not apply to IVDs covered by the IVDR (MDR, Article 1(6)(a)), and an integral product combining both is governed by the MDR, with IVDR requirements for its IVD part. Once a product is under the MDR, its class comes from MDR Annex VIII, which our guide to MDCG 2021-24 walks through.
| Criterion | MDR, Regulation (EU) 2017/745 | IVDR, Regulation (EU) 2017/746 |
|---|---|---|
| Applies from | 26 May 2021 | 26 May 2022 |
| Defined in | Article 2(1), devices for specific medical purposes, including software | Article 2(2), devices examining human specimens in vitro, including software |
| Risk classes | I, IIa, IIb, III (Article 51) | A, B, C, D (Article 47) |
| Classification rules | Annex VIII, Rules 1 to 22 | Annex VIII, Rules 1 to 7 |
| Evidence | clinical evaluation based on clinical data (Article 61) | performance evaluation of scientific validity, analytical performance and clinical performance (Article 56) |
| Notified body | classes IIa, IIb and III, and class I devices that are sterile, have a measuring function or are reusable surgical instruments (Article 52). Custom-made and investigational devices follow their own rules | classes B, C and D, and sterile class A, where it looks only at sterility (Article 48) |
The table draws on the MDR and the IVDR, articles as cited, and takes the application dates from the European Commission overview.
Decide from the intended purpose, in a practical order. Write the purpose down from everything you publish, check whether a human specimen is examined outside the body for one of the IVD purposes, rule out the exclusions, check for an integral medical device, and treat software by where its data come from. Record the reasoning in the technical documentation.
Borderline products are settled by the same test, applied to the whole product and to each part. Where is the specimen examined, do the parts form one object, and what do the claims say? These examples come from MDCG guidance and the borderline manual, and the outcome for your product still depends on your intended purpose.
| Product | Regulation | Why | Source |
|---|---|---|---|
| Breath analyser measuring fractional exhaled nitric oxide (FeNO) | IVDR | Exhaled air has left the body, so it is a gaseous specimen examined outside it. The single-use filter is a class I device under the MDR | Borderline manual v5, April 2026, 2.1.1.1 |
| Swab with integrated reagents | MDR | Medical device and IVD form one object | MDCG 2024-11, 2.6.2 |
| Specimen transport tube with a swab built into the cap | MDR | One object | MDCG 2024-11, 2.6.2 |
| Collection kit with a separate swab and specimen receptacle | swab MDR, receptacle IVDR | Packed together without forming one object, so each part is qualified on its own | MDCG 2024-11, 2.6.1 and 2.6.2 |
| Continuous glucose monitor measuring interstitial fluid in vivo, with a built-in capillary blood measurement | MDR | Integral product | MDCG 2024-11, 2.6.2 |
| Pulse oximeter | MDR | No specimen, as light passes through the fingertip | MDCG 2024-11, 2.3 |
| Lancet, needle, urine collection bag | MDR | Applied to the body to obtain the specimen | MDCG 2024-11, 2.2.2 |
| Software interpreting the optical density from an ELISA reader for diagnosis | IVDR | Turns IVD raw data into diagnostic information | MDCG 2019-11 rev.1, 3.4 |
| Research-use-only reagent | neither | Excluded by Article 1(3)(a), unless intended for in vitro diagnostic examination | MDCG 2024-11, 2.5 |
The FeNO device is a good test of instinct. The patient breathes into it and it looks like respiratory equipment, yet the borderline manual qualifies it as an IVD.
The manufacturer qualifies its product first. The borderline manual records what national competent authorities agreed case by case. It is not legally binding, and decisions stay with national authorities and courts. At a Member State's request, or on its own initiative, the Commission can decide by implementing act whether a product is an IVD (IVDR, Article 3). Classification is a separate question. For a manufacturer established in the EU, a dispute with its notified body over the class goes to the competent authority of its Member State, and a manufacturer with no registered place of business in the EU that has not yet designated an authorised representative goes through the Member State of the representative named in its notified body application (IVDR, Article 47(2)).
IVDR classification sorts IVDs into classes A, B, C and D by intended purpose and inherent risk, under the seven rules of Annex VIII. Apply every rule that fits, and with several rules or stated purposes the higher class wins. The class, plus sterility for class A, decides whether a notified body takes part (IVDR, Articles 47 and 48).
| Rule | Class | What it covers |
|---|---|---|
| 1 | D | transmissible agents in blood, tissue or organ donations, life-threatening transmissible agents with a high or suspected high risk of propagation, and the infectious load of a life-threatening disease where monitoring is critical |
| 2 | C, or D for listed markers | blood grouping and tissue typing for transfusion, transplantation or cell administration, with class D only for the specific ABO, Rhesus, Kell, Kidd and Duffy markers the rule lists |
| 3 | C | among others, sexually transmitted agents, companion diagnostics, cancer screening, diagnosis or staging, human genetic testing, management of life-threatening conditions |
| 4 | C, or B | self-tests are class C, except pregnancy, fertility, cholesterol and urine glucose, erythrocyte, leucocyte and bacteria tests, which are class B. Near-patient tests are classified in their own right |
| 5 | A | general laboratory products, accessories with no critical characteristics, buffer and washing solutions, general culture media and histological stains intended for a specific examination, IVD instruments, and specimen receptacles |
| 6 | B | anything the other rules do not cover |
| 7 | B | controls without a quantitative or qualitative assigned value |
The table follows IVDR, Annex VIII.
Four of the implementing rules come up whenever devices are combined or software is involved. Software that drives or influences a device takes that device's class, and independent software is classified in its own right (1.4). Accessories are classified separately (1.3). Several intended purposes put the device in the higher class (1.8), and so do several applicable rules (1.9). MDCG 2020-16 rev.5 shows how much the wording matters. A test intended to screen blood and tissue donations for syphilis is class D under Rule 1, while a test intended to diagnose syphilis in the individual is class C under Rule 3(a).
The Commission published revision 5 of MDCG 2020-16 on 9 September 2026. It clarifies the rationale of Rule 7 and revises its examples, so a page that still calls revision 4 current is out of date.
Under IVDR Article 48, it depends on the class.
As of 20 November 2026, the IVDR transitional periods of Regulation (EU) 2024/1860 end on 31 December 2027 for IVDD-certified and class D devices, 31 December 2028 for class C and 31 December 2029 for class B and sterile class A. They apply only to legacy devices meeting every condition, and the class D and C application deadlines have passed.
In the Commission's Q&A of July 2024, a legacy device is one of two things. It is either an IVD with a valid certificate issued under the IVDD before 26 May 2022, or an IVD whose IVDD declaration of conformity was drawn up before 26 May 2022 and which now needs a notified body under the IVDR. New devices and non-sterile class A devices have had no transitional period since 26 May 2022.
To stay on the market until those dates, a legacy device had to keep complying with the IVDD, avoid significant changes in design or intended purpose, and pose no unacceptable risk. Its manufacturer had to run an IVDR quality management system from 26 May 2025, apply to a notified body by 26 May 2025 for IVDD-certified and class D devices, 26 May 2026 for class C and 26 May 2027 for class B and sterile class A, and sign a written agreement by 26 September of the same year (Regulation (EU) 2024/1860, amending IVDR Article 110). For previously self-declared class B and sterile class A devices, the application deadline of 26 May 2027 and the agreement deadline of 26 September 2027 are still ahead.
No. The Commission proposed a targeted simplification of the MDR and IVDR on 16 December 2025, COM(2025)1023. On 20 November 2026 the European Parliament's procedure file 2025/0404(COD) shows it awaiting a decision in the Public Health committee, with an indicative plenary sitting on 18 January 2027. It would write the authorities' coordination on qualification and classification, the "Helsinki procedure", into the law, with expert panels able to give opinions. It leaves the transition dates as they are. A new IVDR Article 110(12) would let legacy orphan devices confirmed by an expert panel stay on the market beyond them under conditions, and Article 110(13) would let pending conformity assessments and earlier certificates continue under the current rules. Until it is adopted and applies, plan on the regulations in force.
An IVD can be placed on the EU market only if it complies with the IVDR. A CE mark and declaration issued under the MDR do not show that. The manufacturer must act on the non-conformity, competent authorities can require corrections, restrictions, withdrawal or recall in proportion to the risk, and national penalties apply on top.
I wrote about this after coming across, in a single week, two unrelated companies announcing a CE mark under the MDR for products that, judged by the claims on their own websites, look like IVDs. My concern rests on four provisions.
A manufacturer that considers, or has reason to believe, that a device it placed on the market does not conform must immediately take the corrective action needed to bring it into conformity, withdraw it or recall it, as appropriate, and inform its distributors and, where applicable, its authorised representative and importers. Where the device presents a serious risk, it must also immediately inform the competent authorities of the Member States where it made the device available and, where applicable, the notified body (IVDR, Article 10(11)).
A competent authority that has reason to believe a device is non-compliant evaluates it (Article 89). If the device presents an unacceptable risk, the authority requires corrective action and, in proportion to the risk, restriction, withdrawal or recall within a set period (Article 90). If it does not, the economic operator gets a clearly defined, proportionate period to end the non-compliance, and only if that fails does the Member State restrict, prohibit, withdraw or recall the product (Article 92). So withdrawal is a possible outcome, not an automatic one.
Member States set the penalties, which must be effective, proportionate and dissuasive (IVDR, Article 106, MDR, Article 113). The regulations set no EU-wide fine schedule, so check the law of each Member State where you sell. In Spain, placing on the market products that have not passed the conformity assessment procedures is a very serious infringement, and so is misusing the CE marking on such products. At the top grade the fine runs from €600,001 up to €1,000,000 and may rise to five times the value of the products concerned. For medical devices the top grade applies only where the infringement caused direct harm or a serious and direct risk to public health (Real Decreto Legislativo 1/2015, Articles 112(2)(c) and 114, consolidated text of 31 July 2026). Fines above €1 million are therefore possible, in Spain at least. They are maximums for specific infringements, not a fixed price of a wrong route.
Act in proportion and in order. Confirm the qualification, stop placing new units on the market, assess the risk of units already supplied, inform the supply chain, decide on correction, withdrawal or recall, and then rebuild conformity under the IVDR. Existing evidence can be reused where it is sufficient.
Compared with the five-step list I published on LinkedIn (identify, inform the authority, stop sales, withdraw, rebuild), this order keeps the risk assessment in front of the field action. Not every case needs a withdrawal, and not every document needs rebuilding.
This is how I read the MDR and IVDR as of November 2026. It is general information, not advice on your own product. You qualify it, your notified body assesses that as part of conformity assessment, and your competent authority has the final word, so take your reasoning to them before you build on it.
Nataliya Kuras is CEO and co-founder of Nomius, with 15+ years in regulatory affairs and quality across IVDs, SaMD and AI as a medical device, medical devices and drug-device combination products. She holds RAC-Devices and RAC-Drugs (RAPS) and is an ISO 13485 and ISO 9001 lead auditor.
IVDR stands for the In Vitro Diagnostic Medical Devices Regulation, Regulation (EU) 2017/746. It governs the placing on the EU market of in vitro diagnostic devices, such as reagents, analysers, self-tests and diagnostic software, and has applied since 26 May 2022, when it replaced Directive 98/79/EC.
An IVD, or in vitro diagnostic medical device, is a reagent, kit, instrument, software or similar product that its manufacturer intends to be used outside the body to examine human specimens such as blood, urine or tissue and give medical information. Pregnancy tests, blood glucose meters and laboratory analysers are examples. The IVDR, Regulation (EU) 2017/746, is the EU law that sets what an IVD must meet.
Yes. Software is an IVD when it gives information within IVDR Article 2(2) based on data from IVDs, for example software turning ELISA reader results into a diagnosis. Under MDCG 2019-11 rev.1, software combining IVD and other device data follows whichever data substantially drive its intended purpose.
Non-sterile class A IVDs do not. The manufacturer draws up the technical documentation and signs the EU declaration of conformity itself. Sterile class A IVDs need a notified body, but only for the aspects of establishing, securing and maintaining sterile conditions (IVDR Article 48(10)). Classes B, C and D always need one, and the higher the class, the more of the documentation it assesses.
CE-IVD is informal shorthand for an in vitro diagnostic device that carries the CE marking, not a separate approval. The mark itself does not show the route (IVDR Article 18). The EU declaration of conformity tells you whether it rests on the IVDR (Article 17) or, for a legacy device still in its transition, on the IVDD, so ask for the declaration when the route matters.
It does not show conformity with the regulation that applies. The MDR excludes IVDs covered by the IVDR, and outside the legacy transition an IVD may be placed on the market only after the IVDR conformity assessment its class requires. The manufacturer must take corrective action, and authorities can require more in proportion to the risk.
Yes, as a maximum, in Spain at least. There, placing products on the market without the required conformity assessment is a very serious infringement. Its top grade, a fine of up to €1,000,000 that may rise to five times the value of the products, applies to medical devices only after direct harm or a serious and direct risk to public health.
No, not by default. The IVDR performance evaluation needs scientific validity, analytical performance and clinical performance. MDCG 2022-2 accepts existing data such as literature and earlier studies where they are sufficient, and asks for a gap analysis to decide which new studies are needed for the intended purpose.
The IVDD, Directive 98/79/EC, had to be transposed into each Member State's law, and only devices on its Annex II lists and devices for self-testing involved a notified body. The IVDR, Regulation (EU) 2017/746, has applied directly in every Member State since 26 May 2022, sorts IVDs into classes A to D under seven rules, and needs a notified body for every class except non-sterile class A.
Not as of 20 November 2026. The Commission's proposal COM(2025)1023 of 16 December 2025 is still awaiting a committee decision in the European Parliament, with an indicative plenary sitting on 18 January 2027. The transitional periods of Regulation (EU) 2024/1860 remain the law in force, and the proposal leaves those dates as they are.
How to classify a medical device under EU MDR Annex VIII, every change in MDCG 2021-24 Rev.1 (April 2026), Rule 11 for software, and what the 2025 proposal would change.