Medical device or IVD? MDR IVDR qualification explained

MDR IVDR qualification, settled from the intended purpose before any class, decides which EU regulation applies. A product meant to examine human specimens in vitro for medical information is an IVD under the IVDR, Regulation (EU) 2017/746. Other medical devices fall under the MDR, Regulation (EU) 2017/745. A CE mark under the wrong one shows no conformity.

  • MDR
  • IVDR
  • Qualification
  • Classification

The IVDR is the EU regulation for in vitro diagnostic medical devices. IVDR rules have applied since 26 May 2022, when they replaced Directive 98/79/EC (IVDR, Articles 112 and 113).

To decide, read everything you say about the product against the IVD definition in IVDR Article 2(2) and the device definition in MDR Article 2(1), check the exclusions and the integral-product rule, and only then classify.

Five points to take from this guide

  • Qualification comes before classification, and your website, press releases and sales decks are part of the intended purpose (IVDR, Article 2(12)).
  • A specimen examined outside the body points to the IVDR. A measurement on or in the body, or a device that is invasive or applied directly to the body to obtain the specimen, points to the MDR, and so does an IVD built into a medical device as one object, though its IVD part must still meet IVDR requirements (MDCG 2024-11, October 2024).
  • IVDR classes run from A to D under seven rules, the higher class wins when several apply, and every class except non-sterile class A needs a notified body (IVDR, Article 48 and Annex VIII, with MDCG 2020-16 rev.5, September 2026).
  • An IVD placed on the market under the MDR cannot borrow the IVDR transitional periods. They cover only devices with an IVDD certificate or IVDD declaration of conformity from before 26 May 2022 (Commission Q&A on Regulation (EU) 2024/1860, July 2024).
  • Authorities answer non-compliance in proportion to the risk, and penalties are national. In Spain, placing on the market a product that skipped its conformity assessment can be fined up to €1,000,000 and beyond (Real Decreto Legislativo 1/2015, Articles 112 and 114).
MDR and IVDR dates at a glance, as of 20 November 2026
MilestoneDateApplies toSource
MDR applies26 May 2021medical devicesEuropean Commission, overview
IVDR applies26 May 2022all IVDs, with no transition for new IVDs or non-sterile class AIVDR, Article 113(2) and Commission Q&A, July 2024
IVDR quality management system in place26 May 2025 (passed)legacy IVDs using the transitionRegulation (EU) 2024/1860 and Commission Q&A
Notified body application, then signed agreement26 May 2025, then 26 September 2025 (both passed)IVDD-certified legacy devices and class DCommission Q&A
Notified body application, then signed agreement26 May 2026, then 26 September 2026 (both passed)class C legacy devicesCommission Q&A
Notified body application, then signed agreement26 May 2027, then 26 September 2027class B and sterile class A legacy devicesCommission Q&A
End of the transitional period31 December 2027 / 31 December 2028 / 31 December 2029IVDD-certified and class D / class C / class B and sterile class ACommission Q&A
Commission proposal COM(2025)1023 to simplify MDR and IVDRproposed 16 December 2025, awaiting committee decision, indicative plenary 18 January 2027not lawEuropean Parliament, procedure 2025/0404(COD)

What is IVDR?

IVDR is the In Vitro Diagnostic Medical Devices Regulation, Regulation (EU) 2017/746. It sets the rules for placing in vitro diagnostic devices on the EU market and has applied since 26 May 2022, when it replaced the In Vitro Diagnostic Medical Devices Directive 98/79/EC (IVDD). An IVD is the product, and the IVDR is the law that governs it.

Under IVDR Article 2(2), an IVD is a medical device that is a reagent, calibrator, control material, kit, instrument, apparatus, piece of equipment, software or system, intended by its manufacturer to be used in vitro to examine specimens derived from the human body, including blood and tissue donations. Its purpose, solely or principally, is information on a physiological or pathological process or state, on congenital impairments, on predisposition to a disease, on safety and compatibility with potential recipients, on likely treatment response, or for defining or monitoring therapeutic measures. Specimen receptacles count as IVDs as well.

IVDR Article 1(3) takes four groups out of scope. The first is products for general laboratory use and research-use-only products, unless the manufacturer specifically intends them for in vitro diagnostic examination. The second is invasive sampling products and products applied directly to the body to obtain a specimen. The last two are internationally certified reference materials and materials for external quality assessment schemes.

An IVD placed on the market through the IVDR conformity assessment carries the CE marking (IVDR, Article 18), followed by the notified body's number where one took part. "CE-IVD" is informal shorthand for a CE-marked IVD, not a separate approval, and it does not tell you the route on its own. Legacy IVDs still in their transitional period carry CE marking based on IVDD conformity, and the EU declaration of conformity says which regulation it rests on.

What is the difference between MDR and IVDR?

The MDR covers medical devices in general, including diagnostic devices that work without a specimen. The IVDR covers the medical devices that examine human specimens in vitro. The MDR does not apply to IVDs covered by the IVDR (MDR, Article 1(6)(a)), and an integral product combining both is governed by the MDR, with IVDR requirements for its IVD part. Once a product is under the MDR, its class comes from MDR Annex VIII, which our guide to MDCG 2021-24 walks through.

MDR and IVDR compared on the same criteria
CriterionMDR, Regulation (EU) 2017/745IVDR, Regulation (EU) 2017/746
Applies from26 May 202126 May 2022
Defined inArticle 2(1), devices for specific medical purposes, including softwareArticle 2(2), devices examining human specimens in vitro, including software
Risk classesI, IIa, IIb, III (Article 51)A, B, C, D (Article 47)
Classification rulesAnnex VIII, Rules 1 to 22Annex VIII, Rules 1 to 7
Evidenceclinical evaluation based on clinical data (Article 61)performance evaluation of scientific validity, analytical performance and clinical performance (Article 56)
Notified bodyclasses IIa, IIb and III, and class I devices that are sterile, have a measuring function or are reusable surgical instruments (Article 52). Custom-made and investigational devices follow their own rulesclasses B, C and D, and sterile class A, where it looks only at sterility (Article 48)

The table draws on the MDR and the IVDR, articles as cited, and takes the application dates from the European Commission overview.

How do you decide whether a product falls under MDR or IVDR?

Decide from the intended purpose, in a practical order. Write the purpose down from everything you publish, check whether a human specimen is examined outside the body for one of the IVD purposes, rule out the exclusions, check for an integral medical device, and treat software by where its data come from. Record the reasoning in the technical documentation.

  1. Write down the intended purpose from every source. IVDR Article 2(12) defines it from the label, the instructions for use, promotional or sales materials and statements, and the performance evaluation. A claim on your website is part of it, alongside the label. If even one intended purpose gives information within IVDR Article 2(2), the product must comply with the IVDR. MDCG 2024-11 gives the example of a lay ancestry kit that also reports predisposition to disease.
  2. Is a specimen derived from the human body examined outside it? If no specimen is involved, or the examination happens in or on the body, the product is not an IVD. MDCG 2024-11 places a pulse oximeter, non-invasive glucose measurement by near-infrared energy and an MRI scanner under the MDR.
  3. Does the information fall within IVDR Article 2(2)(a) to (f)? Those points cover a physiological or pathological state, congenital impairment, predisposition, compatibility with recipients, treatment response and therapeutic monitoring.
  4. Does an exclusion in Article 1(3) apply? A product intended for research use only cannot also have a medical purpose, and swabs, lancets and needles that are invasive or applied directly to the body to take the specimen are medical devices under the MDR (MDCG 2024-11, sections 2.2 and 2.5).
  5. Is a medical device built in as an integral part? Where a medical device and an IVD are physically combined into a single object, the product is governed by the MDR, and the IVDR's requirements apply to the IVD part (MDR, Article 1(7), IVDR, Article 1(4)). Parts packed together without forming one object are qualified one by one.
  6. For software, follow the data. Under MDCG 2019-11 rev.1, June 2025, first check that the software is medical device software at all, because storage, transfer or simple search of IVD results is not enough. Software that gives information within IVDR Article 2(2) from IVD data only is then an IVD, and for mixed inputs the regulation follows the data sources that substantially drive the intended purpose. Software that drives or influences a hardware device follows that device's regulation.

Borderline cases between device and IVD

Borderline products are settled by the same test, applied to the whole product and to each part. Where is the specimen examined, do the parts form one object, and what do the claims say? These examples come from MDCG guidance and the borderline manual, and the outcome for your product still depends on your intended purpose.

Borderline products and the regulation that applies, from MDCG guidance and the borderline manual
ProductRegulationWhySource
Breath analyser measuring fractional exhaled nitric oxide (FeNO)IVDRExhaled air has left the body, so it is a gaseous specimen examined outside it. The single-use filter is a class I device under the MDRBorderline manual v5, April 2026, 2.1.1.1
Swab with integrated reagentsMDRMedical device and IVD form one objectMDCG 2024-11, 2.6.2
Specimen transport tube with a swab built into the capMDROne objectMDCG 2024-11, 2.6.2
Collection kit with a separate swab and specimen receptacleswab MDR, receptacle IVDRPacked together without forming one object, so each part is qualified on its ownMDCG 2024-11, 2.6.1 and 2.6.2
Continuous glucose monitor measuring interstitial fluid in vivo, with a built-in capillary blood measurementMDRIntegral productMDCG 2024-11, 2.6.2
Pulse oximeterMDRNo specimen, as light passes through the fingertipMDCG 2024-11, 2.3
Lancet, needle, urine collection bagMDRApplied to the body to obtain the specimenMDCG 2024-11, 2.2.2
Software interpreting the optical density from an ELISA reader for diagnosisIVDRTurns IVD raw data into diagnostic informationMDCG 2019-11 rev.1, 3.4
Research-use-only reagentneitherExcluded by Article 1(3)(a), unless intended for in vitro diagnostic examinationMDCG 2024-11, 2.5

The FeNO device is a good test of instinct. The patient breathes into it and it looks like respiratory equipment, yet the borderline manual qualifies it as an IVD.

Who has the final say on qualification?

The manufacturer qualifies its product first. The borderline manual records what national competent authorities agreed case by case. It is not legally binding, and decisions stay with national authorities and courts. At a Member State's request, or on its own initiative, the Commission can decide by implementing act whether a product is an IVD (IVDR, Article 3). Classification is a separate question. For a manufacturer established in the EU, a dispute with its notified body over the class goes to the competent authority of its Member State, and a manufacturer with no registered place of business in the EU that has not yet designated an authorised representative goes through the Member State of the representative named in its notified body application (IVDR, Article 47(2)).

How does IVDR classification work?

IVDR classification sorts IVDs into classes A, B, C and D by intended purpose and inherent risk, under the seven rules of Annex VIII. Apply every rule that fits, and with several rules or stated purposes the higher class wins. The class, plus sterility for class A, decides whether a notified body takes part (IVDR, Articles 47 and 48).

IVDR Annex VIII classification rules and the class each gives
RuleClassWhat it covers
1Dtransmissible agents in blood, tissue or organ donations, life-threatening transmissible agents with a high or suspected high risk of propagation, and the infectious load of a life-threatening disease where monitoring is critical
2C, or D for listed markersblood grouping and tissue typing for transfusion, transplantation or cell administration, with class D only for the specific ABO, Rhesus, Kell, Kidd and Duffy markers the rule lists
3Camong others, sexually transmitted agents, companion diagnostics, cancer screening, diagnosis or staging, human genetic testing, management of life-threatening conditions
4C, or Bself-tests are class C, except pregnancy, fertility, cholesterol and urine glucose, erythrocyte, leucocyte and bacteria tests, which are class B. Near-patient tests are classified in their own right
5Ageneral laboratory products, accessories with no critical characteristics, buffer and washing solutions, general culture media and histological stains intended for a specific examination, IVD instruments, and specimen receptacles
6Banything the other rules do not cover
7Bcontrols without a quantitative or qualitative assigned value

The table follows IVDR, Annex VIII.

Four of the implementing rules come up whenever devices are combined or software is involved. Software that drives or influences a device takes that device's class, and independent software is classified in its own right (1.4). Accessories are classified separately (1.3). Several intended purposes put the device in the higher class (1.8), and so do several applicable rules (1.9). MDCG 2020-16 rev.5 shows how much the wording matters. A test intended to screen blood and tissue donations for syphilis is class D under Rule 1, while a test intended to diagnose syphilis in the individual is class C under Rule 3(a).

The Commission published revision 5 of MDCG 2020-16 on 9 September 2026. It clarifies the rationale of Rule 7 and revises its examples, so a page that still calls revision 4 current is out of date.

Which IVDs need a notified body?

Under IVDR Article 48, it depends on the class.

  • For non-sterile class A, the manufacturer draws up the technical documentation and the EU declaration of conformity itself.
  • For sterile class A, a notified body takes part, limited to establishing, securing and maintaining sterile conditions.
  • Classes B, C and D need a notified body. For class B it assesses the technical documentation of at least one representative device per category of devices, for class C at least one per generic device group, and class D adds the technical documentation assessment of Annex IX Chapter II. For class D devices where EU reference laboratories are designated, the notified body asks one of them to verify the claimed performance by laboratory testing.

What are the IVDR transition deadlines in November 2026?

As of 20 November 2026, the IVDR transitional periods of Regulation (EU) 2024/1860 end on 31 December 2027 for IVDD-certified and class D devices, 31 December 2028 for class C and 31 December 2029 for class B and sterile class A. They apply only to legacy devices meeting every condition, and the class D and C application deadlines have passed.

In the Commission's Q&A of July 2024, a legacy device is one of two things. It is either an IVD with a valid certificate issued under the IVDD before 26 May 2022, or an IVD whose IVDD declaration of conformity was drawn up before 26 May 2022 and which now needs a notified body under the IVDR. New devices and non-sterile class A devices have had no transitional period since 26 May 2022.

To stay on the market until those dates, a legacy device had to keep complying with the IVDD, avoid significant changes in design or intended purpose, and pose no unacceptable risk. Its manufacturer had to run an IVDR quality management system from 26 May 2025, apply to a notified body by 26 May 2025 for IVDD-certified and class D devices, 26 May 2026 for class C and 26 May 2027 for class B and sterile class A, and sign a written agreement by 26 September of the same year (Regulation (EU) 2024/1860, amending IVDR Article 110). For previously self-declared class B and sterile class A devices, the application deadline of 26 May 2027 and the agreement deadline of 26 September 2027 are still ahead.

Does the December 2025 proposal change these dates?

No. The Commission proposed a targeted simplification of the MDR and IVDR on 16 December 2025, COM(2025)1023. On 20 November 2026 the European Parliament's procedure file 2025/0404(COD) shows it awaiting a decision in the Public Health committee, with an indicative plenary sitting on 18 January 2027. It would write the authorities' coordination on qualification and classification, the "Helsinki procedure", into the law, with expert panels able to give opinions. It leaves the transition dates as they are. A new IVDR Article 110(12) would let legacy orphan devices confirmed by an expert panel stay on the market beyond them under conditions, and Article 110(13) would let pending conformity assessments and earlier certificates continue under the current rules. Until it is adopted and applies, plan on the regulations in force.

What happens if a product was CE marked under MDR but is an IVD?

An IVD can be placed on the EU market only if it complies with the IVDR. A CE mark and declaration issued under the MDR do not show that. The manufacturer must act on the non-conformity, competent authorities can require corrections, restrictions, withdrawal or recall in proportion to the risk, and national penalties apply on top.

I wrote about this after coming across, in a single week, two unrelated companies announcing a CE mark under the MDR for products that, judged by the claims on their own websites, look like IVDs. My concern rests on four provisions.

  • The MDR does not apply to IVDs covered by the IVDR (MDR, Article 1(6)(a)).
  • An IVD may be placed on the market only if it complies with the IVDR (IVDR, Article 5(1)), after the conformity assessment its class requires (Article 48(1)).
  • The EU declaration of conformity states that the IVDR's requirements are met, and the CE marking goes on devices considered to conform with the IVDR (Articles 17(1) and 18(1)). An MDR declaration states neither.
  • An MDR certificate cannot make a product a legacy device. The transitional periods cover only IVDs with an IVDD certificate or an IVDD declaration from before 26 May 2022 (Commission Q&A). Without that earlier IVDD coverage, an IVD placed on the market under the MDR is, for the IVDR, a new device.

What the manufacturer must do

A manufacturer that considers, or has reason to believe, that a device it placed on the market does not conform must immediately take the corrective action needed to bring it into conformity, withdraw it or recall it, as appropriate, and inform its distributors and, where applicable, its authorised representative and importers. Where the device presents a serious risk, it must also immediately inform the competent authorities of the Member States where it made the device available and, where applicable, the notified body (IVDR, Article 10(11)).

What the authorities can do

A competent authority that has reason to believe a device is non-compliant evaluates it (Article 89). If the device presents an unacceptable risk, the authority requires corrective action and, in proportion to the risk, restriction, withdrawal or recall within a set period (Article 90). If it does not, the economic operator gets a clearly defined, proportionate period to end the non-compliance, and only if that fails does the Member State restrict, prohibit, withdraw or recall the product (Article 92). So withdrawal is a possible outcome, not an automatic one.

Penalties for placing a device under the wrong regulation

Member States set the penalties, which must be effective, proportionate and dissuasive (IVDR, Article 106, MDR, Article 113). The regulations set no EU-wide fine schedule, so check the law of each Member State where you sell. In Spain, placing on the market products that have not passed the conformity assessment procedures is a very serious infringement, and so is misusing the CE marking on such products. At the top grade the fine runs from €600,001 up to €1,000,000 and may rise to five times the value of the products concerned. For medical devices the top grade applies only where the infringement caused direct harm or a serious and direct risk to public health (Real Decreto Legislativo 1/2015, Articles 112(2)(c) and 114, consolidated text of 31 July 2026). Fines above €1 million are therefore possible, in Spain at least. They are maximums for specific infringements, not a fixed price of a wrong route.

What to do when you find the product is under the wrong regulation

Act in proportion and in order. Confirm the qualification, stop placing new units on the market, assess the risk of units already supplied, inform the supply chain, decide on correction, withdrawal or recall, and then rebuild conformity under the IVDR. Existing evidence can be reused where it is sufficient.

  1. Confirm the qualification on paper. Test the intended purpose as you actually market it against IVDR Article 2(2), with MDCG 2024-11, MDCG 2019-11 rev.1 and the borderline manual. If the answer is not clear-cut, get an independent regulatory opinion before you act on it.
  2. Stop placing new units on the market. Under IVDR Article 5(1) a non-conforming IVD cannot be placed on the market, so hold release while the route is fixed.
  3. Assess the risk of units already supplied. If they present a serious risk, immediately inform the competent authorities of the Member States where you made them available and, where applicable, the notified body (Article 10(11)).
  4. Decide the field action from that assessment and record it. Choose correction, withdrawal or recall as appropriate, and document the decision in your corrective and preventive action process.
  5. Inform your distributors, and your authorised representative and importers where you have them, of the action (Article 10(11)).
  6. Talk to the competent authority early, even without a serious risk. The law requires notification for serious risk, and my advice is to bring the facts and your plan to your competent authority before it comes to you.
  7. Classify under IVDR Annex VIII and choose the conformity route. For classes B, C, D and sterile class A, a notified body must complete its conformity assessment before you can affix the CE marking.
  8. Run a gap analysis of the technical documentation and the evidence. An MDR clinical evaluation is built on clinical data. The IVDR asks for scientific validity, analytical performance and clinical performance (Article 56). MDCG 2022-2 allows existing data, such as literature and earlier studies, where it is sufficient, and asks for a gap analysis to decide what new evidence to generate. You do not repeat every test by default, and you do not skip what is missing.
  9. Re-enter the market only under the IVDR. That means a completed conformity assessment, an IVDR declaration of conformity and the CE marking with the notified body's number where one took part (Articles 17 and 18).

Compared with the five-step list I published on LinkedIn (identify, inform the authority, stop sales, withdraw, rebuild), this order keeps the risk assessment in front of the field action. Not every case needs a withdrawal, and not every document needs rebuilding.

Common mistakes in MDR and IVDR qualification

  • Classifying before qualifying. A risk class under the wrong regulation answers the wrong question.
  • Reading the intended purpose from the label alone, when website claims, press releases and sales decks count too (Article 2(12)).
  • Treating "diagnostic" as "IVD". A pulse oximeter or an imaging device diagnoses without a specimen and sits under the MDR.
  • Treating a collection kit as one product. A separate swab and receptacle are qualified one by one.
  • Qualifying software from its marketing name alone, or from where its data come from alone, instead of its intended purpose and the MDCG 2019-11 decision steps.
  • Assuming class A means no notified body when the product is sterile.
  • Assuming a transitional period for a device that was never an IVDD legacy device.
  • Presenting MDCG 2020-16 revision 4 as current after revision 5 of September 2026.
  • Planning on the December 2025 proposal as if it were law.
  • Re-running every study after a change of route, or none, instead of a documented gap analysis.

Glossary

  • IVD means an in vitro diagnostic medical device, as defined in IVDR Article 2(2).
  • The intended purpose is the use stated in the label, instructions for use, promotional or sales materials and the performance evaluation (IVDR Article 2(12)).
  • Qualification is deciding whether a product is a medical device, an IVD, an accessory, or neither.
  • Classification is assigning the risk class within the regulation that applies.
  • A specimen receptacle is a device specifically intended for the primary containment and preservation of human specimens for in vitro examination (IVDR Article 2(3)).
  • An integral product is a medical device and an IVD physically combined into a single object. It is governed by the MDR, with the IVDR applying to the IVD part.
  • A legacy device is an IVD covered by an IVDD certificate or IVDD declaration from before 26 May 2022, eligible for the transitional periods if it meets their conditions.
  • A notified body is an organisation designated to assess conformity for the classes that need it.
  • A competent authority is the national regulator that runs market surveillance.
  • Performance evaluation is the IVDR assessment of scientific validity, analytical performance and clinical performance.
  • The Helsinki procedure is the coordination between national competent authorities on qualification and classification questions, which the December 2025 proposal would put into the regulations.

Official resources

  • Regulation (EU) 2017/746 (IVDR), EUR-Lex
  • Regulation (EU) 2017/745 (MDR), EUR-Lex
  • Regulation (EU) 2024/1860, EUR-Lex
  • MDCG guidance index, European Commission
  • MDCG 2024-11, qualification of IVDs
  • MDCG 2020-16 rev.5, IVD classification rules
  • MDCG 2019-11 rev.1, software qualification and classification
  • Manual on borderline and classification, version 5
  • Commission Q&A on the IVDR transition extension
  • HPRA, qualification and classification of IVDs

This is how I read the MDR and IVDR as of November 2026. It is general information, not advice on your own product. You qualify it, your notified body assesses that as part of conformity assessment, and your competent authority has the final word, so take your reasoning to them before you build on it.

Nataliya Kuras is CEO and co-founder of Nomius, with 15+ years in regulatory affairs and quality across IVDs, SaMD and AI as a medical device, medical devices and drug-device combination products. She holds RAC-Devices and RAC-Drugs (RAPS) and is an ISO 13485 and ISO 9001 lead auditor.

Sources

  • Regulation (EU) 2017/746 (IVDR), Articles 1, 2, 3, 5, 10, 17, 18, 47, 48, 56, 89, 90, 92, 106, 112, 113 and Annex VIII, OJ L 117, 5 May 2017, https://eur-lex.europa.eu/eli/reg/2017/746/oj (text read through the EU Publications Office, 20 November 2026)
  • Regulation (EU) 2017/745 (MDR), Articles 1, 2, 51, 52, 61, 113 and Annex VIII, OJ L 117, 5 May 2017, https://eur-lex.europa.eu/eli/reg/2017/745/oj
  • Regulation (EU) 2024/1860 of 13 June 2024, in force 9 July 2024, https://eur-lex.europa.eu/eli/reg/2024/1860/oj
  • European Commission, Q&A on the extended IVDR transitional period, July 2024, https://health.ec.europa.eu/document/download/dfd7a1c6-f319-4682-9bac-77bef1165818_en?filename=mdr_qna-ext-ivdr.pdf
  • European Commission, medical devices new regulations overview, read 20 November 2026, https://health.ec.europa.eu/medical-devices-new-regulations/overview_en
  • MDCG 2024-11, Guidance on qualification of in vitro diagnostic medical devices, October 2024
  • MDCG 2019-11 rev.1, Qualification and classification of software, June 2025
  • MDCG 2020-16 rev.5, Classification rules for IVDs, September 2026 (published 9 September 2026)
  • MDCG 2022-2, General principles of clinical evidence for IVDs, January 2022
  • Manual on borderline and classification under Regulations (EU) 2017/745 and 2017/746, version 5, April 2026
  • European Commission, COM(2025)1023 final, 16 December 2025, and European Parliament Legislative Observatory, procedure 2025/0404(COD), read 20 November 2026
  • Spain, Real Decreto Legislativo 1/2015, Articles 112 and 114, consolidated text updated 31 July 2026, https://www.boe.es/buscar/act.php?id=BOE-A-2015-8343

Frequently asked questions

What does IVDR stand for?

IVDR stands for the In Vitro Diagnostic Medical Devices Regulation, Regulation (EU) 2017/746. It governs the placing on the EU market of in vitro diagnostic devices, such as reagents, analysers, self-tests and diagnostic software, and has applied since 26 May 2022, when it replaced Directive 98/79/EC.

What is an IVD under the IVDR?

An IVD, or in vitro diagnostic medical device, is a reagent, kit, instrument, software or similar product that its manufacturer intends to be used outside the body to examine human specimens such as blood, urine or tissue and give medical information. Pregnancy tests, blood glucose meters and laboratory analysers are examples. The IVDR, Regulation (EU) 2017/746, is the EU law that sets what an IVD must meet.

Can software be an IVD under the IVDR?

Yes. Software is an IVD when it gives information within IVDR Article 2(2) based on data from IVDs, for example software turning ELISA reader results into a diagnosis. Under MDCG 2019-11 rev.1, software combining IVD and other device data follows whichever data substantially drive its intended purpose.

Do class A IVDs need a notified body?

Non-sterile class A IVDs do not. The manufacturer draws up the technical documentation and signs the EU declaration of conformity itself. Sterile class A IVDs need a notified body, but only for the aspects of establishing, securing and maintaining sterile conditions (IVDR Article 48(10)). Classes B, C and D always need one, and the higher the class, the more of the documentation it assesses.

What does CE-IVD mean?

CE-IVD is informal shorthand for an in vitro diagnostic device that carries the CE marking, not a separate approval. The mark itself does not show the route (IVDR Article 18). The EU declaration of conformity tells you whether it rests on the IVDR (Article 17) or, for a legacy device still in its transition, on the IVDD, so ask for the declaration when the route matters.

Is a CE mark under the MDR valid for a product that is an IVD?

It does not show conformity with the regulation that applies. The MDR excludes IVDs covered by the IVDR, and outside the legacy transition an IVD may be placed on the market only after the IVDR conformity assessment its class requires. The manufacturer must take corrective action, and authorities can require more in proportion to the risk.

Can fines for an IVD placed on the market under the wrong regulation exceed €1 million?

Yes, as a maximum, in Spain at least. There, placing products on the market without the required conformity assessment is a very serious infringement. Its top grade, a fine of up to €1,000,000 that may rise to five times the value of the products, applies to medical devices only after direct harm or a serious and direct risk to public health.

Do we have to repeat all testing when moving from the MDR to the IVDR?

No, not by default. The IVDR performance evaluation needs scientific validity, analytical performance and clinical performance. MDCG 2022-2 accepts existing data such as literature and earlier studies where they are sufficient, and asks for a gap analysis to decide which new studies are needed for the intended purpose.

What changed for an IVD when the IVDR replaced the IVDD?

The IVDD, Directive 98/79/EC, had to be transposed into each Member State's law, and only devices on its Annex II lists and devices for self-testing involved a notified body. The IVDR, Regulation (EU) 2017/746, has applied directly in every Member State since 26 May 2022, sorts IVDs into classes A to D under seven rules, and needs a notified body for every class except non-sterile class A.

Has the December 2025 proposal changed the IVDR deadlines?

Not as of 20 November 2026. The Commission's proposal COM(2025)1023 of 16 December 2025 is still awaiting a committee decision in the European Parliament, with an indicative plenary sitting on 18 January 2027. The transitional periods of Regulation (EU) 2024/1860 remain the law in force, and the proposal leaves those dates as they are.

Join the Design Partner Programme

Tell us about your product and the markets it is going into, and we will take you through the programme.

Contact us

Tell Us About Your Compliance Needs

Send us a message with any questions and our team will get back to you.

0 / 4000

Thank you. We will get back to you shortly.

We could not send that. Please check the fields and send again with JavaScript on, or email hello@nomius.io.