MDR vs IVDR, medical device or IVD?
How to decide whether a product falls under the MDR or the IVDR, how IVDR classification and deadlines work as of November 2026, and what to do after a CE mark under the wrong regulation.
MDCG 2021-24 is the Medical Device Coordination Group's guidance on EU medical device classification under the MDR, Regulation (EU) 2017/745. It walks through the definitions, implementing rules and 22 classification rules of MDR Annex VIII. The current version is MDCG 2021-24 Rev.1, published on 20 April 2026. It is guidance, not law, and the class always comes from the MDR.
You classify a device by fixing its intended purpose, checking every Annex VIII rule that could apply, and taking the highest class any of them gives.
| Item | Value | Source and date |
|---|---|---|
| Full title | MDCG 2021-24 Rev.1, Guidance on classification of medical devices | MDCG 2021-24 Rev.1, April 2026 |
| Current version | Rev.1, published 20 April 2026 | Commission announcement, 20 April 2026 |
| First version | Published 4 October 2021 | Commission announcement, 4 October 2021 |
| Legal status | Endorsed by the MDCG (MDR Article 103), not legally binding | Rev.1 disclaimer, April 2026 |
| What it interprets | MDR Article 51 and Annex VIII, with four classes, 7 implementing rules and 22 classification rules | MDR, OJ L 117, 5 May 2017, with the 2019 corrigendum |
| Companion guidance for software | MDCG 2019-11 Rev.1 | MDCG 2019-11 Rev.1, June 2025 |
| Pending change to the rules | COM(2025) 1023, procedure 2025/0404(COD), awaiting committee decision | Legislative Observatory, read 20 November 2026 |
MDCG 2021-24 is the guidance document that explains how to apply the classification rules of the EU MDR to a medical device. It collects the definitions that drive classification, explains each of the 22 rules in Annex VIII, and gives tables of example devices for each rule and most of its sub-rules. The current text is Rev.1 of April 2026, 59 pages long, which replaced the first version of 4 October 2021.
The document covers medical devices and Annex XVI products under the MDR only. For some active Annex XVI products, such as lasers, intense pulsed light equipment and brain stimulation equipment, the class is set by Implementing Regulation (EU) 2022/2347 by way of derogation from Annex VIII. In vitro diagnostic devices are classified under a different regulation, the IVDR, with its own rules. If you are not yet sure which regulation your product falls under, settle that first. Our article on MDR/IVDR qualification covers that decision.
Classification matters because the class decides the conformity assessment route, whether and how far a notified body assesses the technical documentation, and several obligations after the device is on the market. Section 2 of MDCG 2021-24 Rev.1 lists them. Clinical investigation, post-market reporting and UDI traceability depend on the class, and some duties, such as the implant card, depend on whether the device is implantable.
No. MDCG 2021-24 is endorsed by the Medical Device Coordination Group established by Article 103 of the MDR, and its own first page says it is not a European Commission document and that its views are not legally binding. Only the Court of Justice of the EU can give binding interpretations of Union law (MDCG 2021-24 Rev.1, April 2026).
In practice it carries weight. It is the agreed reading of the Member States' representatives, a notified body reviewing your class will compare your rationale with it, and Annex II, Section 1.1(f) of the MDR requires your technical documentation to state the risk class and justify the classification rules applied. A rationale that departs from the guidance can still be right, but it has to argue from the wording of Annex VIII, and you should expect the question.
Rev.1 is a targeted revision. Its change table on page 3 lists fourteen updated items (MDCG 2021-24 Rev.1, April 2026), and none of them changes the MDR itself. Annex VIII has not been amended since the 2017 text, apart from a 2019 corrigendum to implementing rule 3.2. What moved are the guidance's explanations, example tables and practical notes. The table below takes every entry in the change table and shows, from a line-by-line comparison of the April 2026 and October 2021 texts, what it changes in practice.
| Where | What Rev.1 changed | Who should re-check |
|---|---|---|
| 2.1 General requirements | Two sentences added. The guidance may be relevant to Annex XVI products except where it is specific to a medical purpose, and it is read together with the guidance written for Annex XVI products | Makers of Annex XVI products without a medical purpose |
| 3.1.4 Implantable devices | The explanation of 'procedure' now says 'clinical procedure' instead of 'surgical procedure', and adds that a clinical procedure covers surgical and non-surgical procedures | Long-dwell devices placed without surgery |
| 3.2 Application of the rules | 'Device' replaced by 'product' once in section 3.2 | No change in substance |
| 4.1.3 Rule 9 diagram | Diagram restructured to follow the order of the rule | No change in substance |
| Rule 2 table | Marked as updated, but the text differs from 2021 only in punctuation | None identified |
| Rule 8 table | Hooks that fix rods on the spinal column move from the Class III examples to the Class IIb examples, and the Class III example of stems implantable in contact with the spinal column is removed | Spinal fixation systems |
| Rule 8 notes | Two new notes. Note 1 says the Article 52(4) list does not make those devices Class IIb, and Note 7 defines contact with the spinal column and treats rod-fixing hooks like screws, wedges, plates and instruments. The five earlier notes stay, renumbered | Spinal and orthopaedic implants, and anyone who read Article 52(4) as a class |
| Rule 9 notes | The paragraph on illumination devices for diagnosis is removed from Rule 9 | Ophthalmic and examination devices |
| Rule 10 table and Note 1 | Surgical lamps added as Class I examples, devices for visual electrophysiology and fundoscopes added as Class IIa examples, and Note 1 reworded around purposes that go beyond illumination | Ophthalmic diagnostic devices, surgical lighting |
| Rule 12 table | The Class IIb example now reads 'pressure regulators for medical gases with integrated flowmeter' | Medical gas equipment |
| Rule 16 table | The Class IIa steriliser example now covers sterilisers for invasive or non-invasive devices in a medical environment | Steriliser makers |
| Rule 22 table | AEDs are Class III 'including their pads/electrodes if they are an integral part of the defibrillators' | AED and electrode makers |
| Rule 22 notes | New Note 2 says defibrillator pads or electrodes placed on the market on their own may be considered under Rule 9 or Rule 1 | Makers of stand-alone pads |
Rev.1 does not change the legal definition. Article 2(5) of the MDR has defined an implantable device by 'clinical intervention' since 2017. What Rev.1 changes is the guidance's explanation of the 'procedure' after which an implant stays in place. It now says 'clinical procedure' and states that this covers surgical and non-surgical procedures (MDCG 2021-24 Rev.1, section 3.1.4). If your classification rationale argued that a device was not an implant because it is placed without surgery, re-read it. Duration on its own does not make a device an implant, though, because the 'at least 30 days' condition in Article 2(5) applies to devices that are partially introduced into the body.
Under Rev.1, hooks that fix rods on the spinal column follow the same exception as screws, wedges, plates and instruments in the last indent of Rule 8, so the guidance lists them as Class IIb examples (Rule 8 table and Note 7). In the 2021 version they sat among the Class III examples. Devices placed in the disc space, interbody fusion devices and spinal disc replacements stay Class III. A hook that meets another Class III indent of Rule 8, for example a biological effect, is still Class III under implementing rule 3.5.
No. Article 52(4) of the MDR sets the conformity assessment for devices that are already Class IIb. For Class IIb implants the notified body assesses the technical documentation of every device, except for the types on its list, which are assessed per generic device group instead. Rev.1 Note 1 to Rule 8 says the list does not classify sutures, staples, screws, plates or the other listed types as Class IIb. Each is classified on its own intended purpose (MDCG 2021-24 Rev.1).
The list Note 1 quotes has since changed. Commission Delegated Regulation (EU) 2026/1359 of 20 March 2026, published on 29 June 2026, replaced it from 19 July 2026. It keeps the original types and adds cannulas, catheters, feeding tubes, suture pledgets, suture sleeves, suture buttons, gastrostomy buttons, bone wax, bone fillers, bone substitutes, stem centralisers, diaphyseal obturators, radiography markers, fibre ligatures, transpalatal distractors, nails, anchors, spinal posterior fixations, textile braids, dental implants, orthodontic devices, dental barriers, suspensory fixations and cinches. That changes how a Class IIb implant is assessed, not its class.
Only when they are an integral part of the defibrillator. Rule 22 puts active therapeutic devices with an integrated diagnostic function that significantly determines patient management, such as automated external defibrillators, in Class III (Annex VIII, Rule 22). Rev.1 ties the Class III example to pads and electrodes that are an integral part of the AED, and its new Note 2 says pads placed on the market on their own may be considered under Rule 9 or Rule 1 (MDCG 2021-24 Rev.1). For a separately sold pad this is a clarification, and the outcome may be a lower class than a reading of the 2021 example suggested.
The Rule 11 section on software is word for word the same as in 2021, including its examples (MDCG 2021-24 Rev.1). Software guidance did move elsewhere. MDCG 2019-11 Rev.1, published in June 2025, clarified sub-rule 11a, added examples and reworked the section on modules. The Rule 11 section of this article covers both.
Annex VIII of the MDR has three chapters. Chapter I defines the terms the rules use, among them duration of use, body orifice, surgically invasive device, reusable surgical instrument, active therapeutic device, the central circulatory and central nervous systems, and injured skin or mucous membrane. Chapter II holds seven implementing rules, numbered 3.1 to 3.7, that govern how the rules are applied. Chapter III holds the 22 classification rules, grouped by the kind of device.
Every device is checked against the general rules for its type and against the special rules, which apply whenever their own conditions are met, whatever kind of device it is. MDR Article 51 sets the four classes, and Article 52(7) adds the Class I subgroups that need limited notified body involvement. These are sterile devices (Is), devices with a measuring function (Im) and reusable surgical instruments (Ir).
| Rule | What it covers | Classes |
|---|---|---|
| Rule 1 | Non-invasive devices not covered by another rule | I |
| Rule 2 | Non-invasive devices for channelling or storing blood, body liquids, cells, tissues, liquids or gases for administration | I, IIa, IIb (blood bags) |
| Rule 3 | Non-invasive devices that modify the composition of blood, body liquids or other liquids, and substances used in vitro with cells, tissues, organs or embryos | IIa, IIb, III |
| Rule 4 | Non-invasive devices in contact with injured skin or mucous membrane, and invasive devices in contact with injured mucous membrane | I, IIa, IIb |
| Rule 5 | Invasive devices via body orifices, not surgically invasive | I, IIa, IIb |
| Rule 6 | Surgically invasive devices for transient use | I (reusable surgical instruments), IIa, IIb, III |
| Rule 7 | Surgically invasive devices for short-term use | IIa, IIb, III |
| Rule 8 | Implantable devices and long-term surgically invasive devices | IIa, IIb, III |
| Rule 9 | Active therapeutic devices that administer or exchange energy, and devices controlling them | IIa, IIb, III |
| Rule 10 | Active devices for diagnosis and monitoring, and ionising radiation for radiology | I, IIa, IIb |
| Rule 11 | Software that informs diagnostic or therapeutic decisions, or monitors physiological processes | I, IIa, IIb, III |
| Rule 12 | Active devices that administer or remove medicinal products, body liquids or other substances | IIa, IIb |
| Rule 13 | All other active devices | I |
| Rule 14 | Devices incorporating a medicinal substance with ancillary action | III |
| Rule 15 | Contraception and prevention of sexually transmitted diseases | IIb, III |
| Rule 16 | Disinfecting, cleaning, rinsing or hydrating contact lenses, and disinfecting or sterilising devices | IIa, IIb |
| Rule 17 | Devices for recording diagnostic X-ray images | IIa |
| Rule 18 | Devices made from non-viable human or animal tissues or cells, or their derivatives | III, unless animal-derived and for intact skin only (then the other rules apply) |
| Rule 19 | Devices incorporating or consisting of nanomaterial | IIa, IIb, III |
| Rule 20 | Invasive devices via body orifices that administer medicinal products by inhalation | IIa, IIb |
| Rule 21 | Substance-based devices introduced via a body orifice or applied to the skin and absorbed or locally dispersed | IIa, IIb, III |
| Rule 22 | Active therapeutic devices with an integrated diagnostic function that significantly determines patient management | III |
Rules 1 to 4 cover non-invasive devices, Rules 5 to 8 invasive devices, Rules 9 to 13 active devices, and Rules 14 to 22 are the special rules. Software sits in the active group under Rule 11, not among the special rules.
Annex VIII, Section 1 defines three durations of continuous use. Transient means normally intended for continuous use for less than 60 minutes. Short term means between 60 minutes and 30 days. Long term means more than 30 days. The duration is what the manufacturer intends for the device, not what happens in a particular patient.
Continuous use has its own definition in implementing rule 3.6. A temporary interruption during a procedure, or a temporary removal for cleaning or disinfection, does not stop the clock, and whether an interruption is temporary is judged against the use before and after it. A device the manufacturer intends to be replaced immediately with another of the same type counts as one accumulated use. MDCG 2021-24 gives three examples. A scalpel used for seconds at a time during a long operation is transient, a ureteric catheter replaced immediately with an identical one extends continuous use, and contact lenses taken out overnight for cleaning are counted only for the hours of uninterrupted wear (MDCG 2021-24 Rev.1, section 3.1). For some products the guidance reads duration as duration of effect, so a cream applied in seconds that stays on the skin for hours is judged on the hours.
Duration is one input, not a class. A short-term device is not automatically Class IIa. Under Rule 7 a short-term surgically invasive device can be IIa, IIb or III depending on where it is used and what it does, and the same holds for long-term devices under Rule 8.
The manufacturer classifies the device, and Article 51(1) of the MDR requires it to be done in accordance with Annex VIII, taking into account the intended purpose and inherent risks. The order below follows the MDR and the guidance.
If your notified body disagrees with your class, Article 51(2) sends the dispute to the competent authority of the Member State where you have your registered place of business, or, for a manufacturer with no registered place of business in the EU that has not yet designated an authorised representative, where the representative named in its notified body application is established. The Commission can also decide the classification of a device or group of devices by implementing act, at a Member State's request under Article 51(3) or on its own initiative under Article 51(4).
The highest class wins. Implementing rule 3.5 of Annex VIII says that where several rules, or several sub-rules within one rule, apply to the same device based on its intended purpose, the strictest rule and sub-rule resulting in the higher classification apply. You cannot stop at the first rule that fits, and you cannot pick the rule that gives the lowest class.
Three other implementing rules decide what is being classified. Under 3.2, a device used in combination with another device is classified separately, and an accessory is classified in its own right, separately from the device it is used with. Under 3.3, software that drives or influences the use of a device falls within that device's class, while independent software is classified on its own. Under 3.4, a device not intended for one specific part of the body is classified on its most critical specified use.
Rule 11 of Annex VIII has three parts. Software intended to provide information used to take decisions with diagnostic or therapeutic purposes is Class IIa, or Class IIb if such decisions may cause a serious deterioration of a person's state of health or a surgical intervention, or Class III if they may cause death or an irreversible deterioration. Software intended to monitor physiological processes is Class IIa, or Class IIb for vital physiological parameters whose variations could result in immediate danger to the patient. All other software is Class I.
| Sub-rule and class | Condition | Example in MDCG 2021-24 Rev.1 |
|---|---|---|
| Decisions, Class IIa | Information used for diagnostic or therapeutic decisions | Software that ranks chemotherapy options for a clinician from patient history and imaging results |
| Decisions, Class IIb | Decisions may cause serious deterioration of health or a surgical intervention | An app that analyses a user's heartbeat, detects abnormalities and informs a physician |
| Decisions, Class III | Decisions may cause death or irreversible deterioration of health | Image-analysis software that diagnoses acute stroke for treatment decisions |
| Monitoring, Class IIa | Monitors physiological processes that are not vital, or vital signs in routine check-ups | Readings of vital signs in routine check-ups, home monitoring included |
| Monitoring, Class IIb | Monitors vital parameters whose variations could cause immediate danger | Continuous surveillance of vital processes in anaesthesia, intensive or emergency care |
| All other, Class I | No decision or monitoring purpose of the kinds above | A conception app that calculates fertility status from basal body temperature and cycle data |
Qualification comes before Rule 11. Software used with medical devices that only records, stores or displays information would generally not be a device (MDCG 2021-24 Rev.1, Rule 11). The rest of Annex VIII comes after it. MDCG 2019-11 Rev.1 (June 2025) says that for software Rules 9, 10, 11, 12, 13, 15 and 22 should all be considered, with implementing rule 3.5 deciding between them. Its own example is melanoma image-analysis software that drives a Class IIa near-infrared scanner. Under 3.3 it would take the scanner's class, but Rule 11 applies to its diagnostic purpose and makes it Class III.
The 2025 software changes sit in MDCG 2019-11 Rev.1. It clarifies sub-rule 11a, including software intended to prevent the risk of illness, adds new examples, says more on writing a clear intended purpose, reworks the section on modules and updates the position of electronic health record systems under the European Health Data Space Regulation (MDCG 2019-11 Rev.1, change table). Its Annex III maps the IMDRF risk categories for software to the possible Rule 11 classes as an illustration for sub-rule 11a. It leaves out Class I software, so treat it as a cross-check rather than a classification.
The rows below start from the guidance's own examples. The dressing rows change one fact at a time, and the other rows set contrasting cases side by side. They show how the rules work. They are not a classification of any real product, which always depends on its full intended purpose.
| Device and assumption | Rule applied | Class |
|---|---|---|
| Wound dressing used as a mechanical barrier or to absorb exudate | Rule 4 | I |
| The same dressing, now claimed to manage the micro-environment of the wound | Rule 4 | IIa |
| The same dressing, intended principally for wounds that breach the dermis and heal only by secondary intent | Rule 4 | IIb |
| Pedicle screw or rod-fixing hook on the spine, no other Class III feature | Rule 8, last indent and Rev.1 Note 7 | IIb |
| Interbody fusion device placed in the disc space | Rule 8, spinal column indent | III |
| AED sold with its pads as an integral part | Rule 22 | III |
| Defibrillator pads placed on the market on their own | Rule 9 or Rule 1, per Rev.1 Rule 22 Note 2 | IIa or IIb under Rule 9, or I under Rule 1, depending on intended purpose and design |
| Software that analyses a user's heartbeat and informs a physician of abnormalities | Rule 11, decisions | IIb |
| Diagnostic software that also drives a Class IIa scanner | Rule 11 with implementing rules 3.3 and 3.5 | III in the MDCG 2019-11 melanoma example |
The spinal example also shows how classification and conformity assessment separate. A rod-fixing hook is a Class IIb implant under Rule 8. Whether its technical documentation is assessed device by device or per generic device group depends on whether it falls within a type on the Article 52(4) list as amended in 2026, such as spinal posterior fixations. That is a question for the route, not for the class.
Each of these errors weakens a classification rationale, and each traces to a provision of the MDR or to the guidance.
Article 52 of the MDR sets the route for each class. The higher the class, the more of the technical documentation a notified body assesses, up to every device for Class III devices and for Class IIb implants not on the Article 52(4) list. Custom-made devices follow Annex XIII instead, with Annex IX Chapter I or Annex XI Part A on top for Class III implantable custom-made devices (Article 52(8)), and investigational devices follow Articles 62 to 81. The table summarises the routes for other devices as they stand on 20 November 2026.
| Class | Route under Article 52 | Notified body | Other duties that depend on the class |
|---|---|---|---|
| I | EU declaration of conformity after drawing up the Annex II and III documentation (Art. 52(7)) | None | Post-market surveillance report, updated when necessary (Art. 85) |
| Is, Im, Ir | Annex IX Chapters I and III, or Annex XI Part A (Art. 52(7)) | Limited to sterility, metrology or reuse aspects | As Class I |
| IIa | Annex IX Chapters I and III with technical documentation of at least one representative device per category, or Annex II and III documentation with Annex XI Section 10 or 18 (Art. 52(6)) | Yes | PSUR when necessary and at least every two years (Art. 86) |
| IIb | Annex IX Chapters I and III with technical documentation of at least one representative device per generic device group, or Annex X with Annex XI. For implants not on the amended Article 52(4) list, every device's documentation is assessed (Art. 52(4), DR 2026/1359) | Yes | PSUR at least annually (Art. 86), and the clinical evaluation consultation procedure for active devices that administer or remove a medicinal product under Rule 12 (Art. 54) |
| III | Annex IX, or Annex X with Annex XI (Art. 52(3)) | Yes | PSUR at least annually, a clinical investigation with exceptions (Art. 61(4) and 61(6), the latter amended by DR 2026/1451), a summary of safety and clinical performance (Art. 32), and the clinical evaluation consultation procedure for implantable devices (Art. 54) |
Implantable devices of any class, other than custom-made or investigational ones, also need a summary of safety and clinical performance under Article 32, clinical investigations under Article 61(4) subject to its exceptions, and an implant card under Article 18, with the exemptions that article lists. The list of implantable and Class III device types exempted from clinical investigations in Article 61(6)(b) was widened by Delegated Regulation (EU) 2026/1451, in force from 19 July 2026, the same day as the Article 52(4) change. A listed type is exempt only where its clinical evaluation rests on sufficient clinical data and meets any product-specific common specification, and the clinical evaluation itself is still required. Article 54(2) lists the cases where the clinical evaluation consultation procedure is not required, such as certificate renewals.
Not on 20 November 2026. On 16 December 2025 the Commission proposed COM(2025) 1023, a regulation that would amend the MDR and the IVDR to reduce their burden. As of 20 November 2026 the Legislative Observatory shows procedure 2025/0404(COD) at the stage 'awaiting committee decision' in the Parliament's Public Health committee, with an indicative plenary date of 18 January 2027. Until a text is adopted and applies, Annex VIII as it stands governs classification, and a rationale should not rely on the proposed wording.
| Provision | Proposed change |
|---|---|
| Implementing rule 3.2 | Wording on accessories for Annex XVI products |
| Rule 2 | Blood bags Class IIb 'by derogation from any other classification rule' |
| Rules 6 and 7 | Reusable surgical instruments Class I regardless of the body part they contact |
| Rule 8 | Accessories to active implantable devices no longer named in the Class III indent, and 'well-established technology devices' added to the exceptions for joint replacements and spinal implants |
| Rule 9 | Active Annex XVI products that administer or exchange energy brought into the rule |
| Rule 11 | Rewritten. Software that generates an output conferring a clinical benefit starts at Class I and rises to IIa, IIb or III with the seriousness of the situation and whether the output informs or drives clinical management |
| Rule 19 | Nanomaterial defined by reference to Commission Recommendation C/2022/3689 |
| Rule 21 | Substances 'absorbed by or locally dispersed on or in the human body' |
For software makers the Rule 11 rewrite is the change that matters most, because it would move the starting class from IIa to I. What it means for a given product depends on the final text, which the Parliament and the Council can still amend, so keep the current rationale and note the proposal as a review trigger.
This is how I read MDCG 2021-24 and Annex VIII as of November 2026, and it is general information rather than advice on your device. The class of your device is yours to justify. Where your route involves a notified body, it checks that class, and only a formal disagreement goes to your competent authority, so test your rationale with your notified body before you build on it.
Nataliya Kuras is CEO and co-founder of Nomius and has 15+ years in regulatory affairs and quality across IVDs, SaMD and AI as a medical device, medical devices and drug-device combination products. RAC-Devices and RAC-Drugs (RAPS), ISO 13485 and ISO 9001 lead auditor.
The official MDCG 2021-24 PDF is on the European Commission's public health website, in its list of MDCG endorsed guidance under borderline and classification. Download the Rev.1 file dated April 2026, which the Commission published on 20 April 2026. Copies of the October 2021 version still circulate on third-party sites and miss the fourteen changes that Rev.1 made.
No. MDCG 2021-24 covers medical devices and Annex XVI products under Regulation (EU) 2017/745. IVDs are classified under Regulation (EU) 2017/746, the IVDR, which has seven classification rules of its own and its own MDCG guidance, MDCG 2020-16, now in revision 5 of September 2026. Our article on MDR/IVDR qualification helps you decide which regulation applies first.
No. The Rule 11 section of Rev.1 is word for word the text of October 2021, examples included. The software guidance that changed is MDCG 2019-11 Rev.1 of June 2025. The December 2025 simplification proposal, COM(2025) 1023, would rewrite Rule 11 itself, but on 20 November 2026 it was still waiting for a committee decision in the European Parliament and was not law.
The manufacturer classifies the device under MDR Annex VIII and justifies the class in its technical documentation. Where the route involves a notified body, it checks that classification. A dispute between the two goes to the competent authority of the Member State where the manufacturer has its registered place of business, or, for a manufacturer outside the EU that has not yet designated an authorised representative, where the representative named in its notified body application is (MDR Article 51(2)).
MDR Annex VIII contains 22 classification rules. Rules 1 to 4 cover non-invasive devices, Rules 5 to 8 invasive devices, Rules 9 to 13 active devices including software, and Rules 14 to 22 are special rules. Seven implementing rules in Chapter II of Annex VIII decide how the rules are applied, and where several rules apply, the strictest one giving the higher class wins.
MDCG stands for the Medical Device Coordination Group, which Article 103 of the EU MDR sets up. Each Member State appoints members with expertise in medical devices and in in vitro diagnostics, who represent its competent authority, and a representative of the European Commission chairs the group. The group endorses guidance such as MDCG 2021-24, which sets out its agreed reading of the MDR but is not legally binding.